Medi-Vac Canister System - Atmos C 341 Battery Operating Instructions Manual

Hide thumbs Also See for C 341 Battery:
Table of Contents

Advertisement

3.4.3 Medi-Vac
WARNING
Risk of infection by contaminated canister system and hoses.
Deadly diseases can be transmitted.
Only use Medi-Vac
Only use sterile packaged parts, when the packaging is undamaged.
Wear disposable gloves.
WARNING
No vacuum or vacuum is too low because of incorrect connection.
Patient can suffocate.
Please observe the operating instructions from the manufacturer of the Medi-Vac
canister system.
Removal
1. Remove the disposable suction hose from the hose guide.
2. Remove the disposable suction hose and the angle
bag.
3. Close the connection „patient" at the Medi-Vac
4. Remove the red hose
5. Close the connection „Vacuum" at the Medi-Vac
6. Remove the vacuum hose from the red connection
canister.
7. Remove the Medi-Vac
8. If necessary: Remove the vacuum hose from the device.
26
Setting up and starting up
canister system
®
suction bags with integrated bacterial filter.
®
1
2
3
4
5
from the Medi-Vac
1
canister system from the support.
®
from the Medi-Vac
2
suction bag with the blue cap
®
suction bag.
®
suction bag with the blue cap
®
of the Medi-Vac
5
®
suction
®
.
3
.
4
outer
®