Travel Speed; Environmental Conditions - Zeiss CONTURA Operating Instructions Manual

Coordinate measuring machine
Table of Contents

Advertisement

CMM
Air quality according to ISO 8573-1:2010
[6:4:4]
Air consumption
Size
Creep speed: SLOW function
Set-up mode:
Batch measurement mode, max.:
Axis (X)
Axis (Y)
Axis (Z)
Vector
Environmental operating temperature
Relative humidity
Reference temperature
Height above mean sea level
4-6
– Air purity with regard to particle concentration ac-
cording to class 6
– Air purity with regard to pressure dew point ac-
cording to class 4
– Air purity with regard to oil concentration accord-
ing to class 4
Max. 80 Nl/min at 1 bar

Travel speed

X/Y/6
[mm/s] 0 - 5
[mm/s] 0 - 70
[mm/s] 300
[mm/s] 220
[mm/s] 300
[mm/s] 470

Environmental conditions

The following conditions must be met to properly operate the coordi-
nate measuring machine.
17-35°C
40 - 70 %
18 - 22 °C, freely selectable
18 - 26 °C (HTG), freely selectable
100 - 240 V: max. 5000 m
Reference temperature
Temperature at which the promised measuring accuracy is guaranteed
when the conditions are met.
The following conditions must be met:
– Temperature for measuring operations must be constant for at least
48 hours.
– Temperature variations must not exceed certain values. Limit values:
1.5 K/d, 1 K/h, and 1 K/m.
61211-1310402
X/Y/8
0 - 5
0 - 70
270
200
300
450
CONTURA