Table 2-3
Source of System Dynamic Accuracy Errors:
The factors affecting dynamic accuracy listed below are primarily a function of the IF detector. However, compression is primarily a function of the
sampler/mixer circuitry. In order to measure these values, some of the system cables must be disconnected to gain access to the individual
instruments.
IF Amplifier Gain Accuracy
Detector Circularity Error:
IF Residuals:
IF Linearity:
Incremental Phase Accuracy (Phase versus Phase) at Measurement Reference:
±0.001 degrees/degree, not to exceed 0.02 degrees peak.
Operating Temperature:
Power:
Dimensions:
Weight:
1. Measured at the IF input to the 8510, not at the test set test ports.
Specifications
8510/8511B Characteristics (continued)
IF Amplifier Power Range (dBm)
–10 to –34
–34 to –46
–46 to –58
–58 to –70
≤ 70
±0.003 dB peak
1
–140 dBm
±0.003 dB
0 °C to 55 °C
110, 120, 220 or 240 ±10% Vac; 47 to 66 Hz line frequency
460 mm x 133 mm x 609 mm (18.1 x 5.25 x 24 inches)
13 kg (29 lb) net; 17 kg (38 lb) shipping
2-4 Agilent 8511B Test Set Operating and Service Manual