Installation - Agilent Technologies 8511B Operating And Service Manual

Frequency converter test set
Table of Contents

Advertisement

Operation

Installation

Installation
Set each test set rear panel address switch to the address listed in
Table 4-2
on page 4-7
if using a two test set configuration, and
Table 4-3 on page 4-11
if configuring more than two test sets. Us the supplied test set interconnect
cable to connect test set number 2, J11, to test set number 1, J10. You may
continue this test set "daisy chain" to include up to four test sets if the total
length of all test set interconnect cables does not exceed 13 meters (about 40
feet). The last test set in the chain does not require Option 001.
If the RF coaxial switch(s) is not incorporated into the system, then the RF
input to the test set must be manually switched to the active test set.
4-8 Agilent 8511B Test Set Operating and Service Manual