Operator's Check; Procedure - Agilent Technologies 8511B Operating And Service Manual

Frequency converter test set
Table of Contents

Advertisement

Table 4-1
NOTE

Procedure

NOTE

Operator's Check

The purpose of this check is to confirm that the Agilent 8511B functions
properly as part of an 8510 system.
Necessary Equipment
Description
Network analyzer system
Semi-rigid cables (2)
Power splitter
1. Supplied with the 8511B service kit (part number 08511-60016).
This procedure must be performed with a properly configured and
operational 8510 system.
Plug in and turn on the frequency converter. The line LED should light
immediately and the active LED should light in about two seconds. Those
indications mean that the instrument has passed its self-test. In case of
difficulty, refer to
Chapter 6, "Test Set Troubleshooting"
local Agilent Service office.
Turn on the source, then the test set and then the Agilent 8510.
On the Agilent 85102:
1. Press
[RECALL], [MORE], [FACTORY PRESET]
8510C, or
for the 8510B.
[PRESET]
Agilent 8360 series systems:
Press STIMULUS,
[MENU], [STEP]
sweep time to 200 ms. In narrow band systems, the power level in the
frequency band generated should match the level shown in
page 4-5
for a given frequency.
2. Reduce the source power by pressing STIMULUS,
,
[MENU]
[POWER SOURCE 1]
All of the observed traces should decrease from
35 ±5 dB at 50.0 GHz. Refer to
Agilent Model or Part Number
8510,B,C
08511-20025
08511-20031
1167C
to preset the
and then use the entry keys to set the
,
,
.
[-10]
[x1]
15 ±5 dB at 45 MHz to
Figure 4-4 on page
Agilent 8511B Test Set Operating and Service Manual 4-3
Operation
Operator's Check
1
1
1
or contact your
Figure 4-4 on
, POWER,
[MENU]
4-5.