Cleaning, Analog-Adjustable Model, Fixed-Volume Model - BRAND Dispensette S Operating Manual

Bottle-top dispenser
Hide thumbs Also See for Dispensette S:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 42
prior to dismantling the instrument
prior to autoclaving
For proper cleaning and removal of any deposits in the parts through which liquids pass, also always
completely with draw the piston from the cylinder after rinsing with a suitable cleaning solution. If
necessary, the parts can also be cleaned in an ultrasonic bath.
9.1 Cleaning, Analog-adjustable model, Fixed-volume
model
1. Completely emptying the instrument
a.
Screw the instrument onto an empty bottle and empty it completely by dispensing. If the instru-
ment is equipped with a recirculation valve, it must be emptied in both the 'dispense' and 'recir-
culate' settings.
2. Rinsing the instrument
a.
Screw the instrument onto a bottle filled with a suitable cleaning agent (e.g. deionized water)
and rinse the instrument several times by completely filling and emptying it.
3. Removing pistons
Avoid swapping pistons
The pistons and cylinders are individually matched, and should not be swapped with pistons from
other instruments!
a.
b.
997490
regularly when liquids accumulate in the
screw cap.
NOTICE
a.
Hold the housing securely and unscrew the piston seat
completely by turning it to the left.
b.
Pull out the piston carefully.
9 Cleaning