About This Guide

It provides the instructions to install the Nitrogen Phosphorus Detector (NPD) in
series with the non-destructive Electron Capture Detector (ECD) already installed
on your TRACE GC Ultra.
NOTE
For Conventions, Symbols and Costumer Communications, please refer to the TRACE GC
Ultra Operating Manual.

Introduction

The NPD tandem option (stacked configuration) is particularly useful for
confirmation type analyses permitting collection of two detector signals
originated from the same injection.
A particular zero dead volume heated adapter allows for correct sample transfer
from the ECD to the NPD avoiding any condensation and peak broadening.
The upgrade kit contains all the material required to install the NPD detector in
series with the ECD detector:
The NPD must be installed and configured as auxiliary detector.

Who Performs the Installation of the Kit

If, for any reason, the kit is not installed by ThermoFinnigan technical personnel,
you must carefully adhere to the following instructions.
6
NPD Assembly for TRACE
NPD Control Card
NPD Standard Outfit
Heater Adapter
Transformer for NPD
DGFC Module for NPD
About This Guide
Installation Guide