Thermo Scientific NPD Tandem Upgrade Kit Installation Manual page 22

For trace gc ultra
Table of Contents

Advertisement

Upgrade Kit
NPD Tandem Detector
3. As the Power Control Module is being removed, disconnection of the parts
4. Place the Power Control Module on a free working table.
22
Figure 1-17. TRACE GC Ultra Rear View
must be carried out in the following order:
• J1, 16-pin white connector.
• J8, 2-pin connector.
• J11, 16-pin white connector located in the upper rear quadrant of the
chassis.
• J13, ribbon cable.
• Oven heater wire connected to the oven heater triac.
• TB1 and TB2, orange terminal strips for the heater wire connections.
• J17, oven blower motor.
• J8, second oven heater wire.
• Chassis ground wire from the GC rear oven terminal.
1