Getting Started

Install the Transformer for Detector

Figure 1-18 shows the positions on the GC Power Control Module where the
detector transformer must be installed.
Install the transformer proceeding as follows. To identify the parts, refer to
Figure 1-19 on page 24.
1. Screw the spacer on the threaded pin.
2. Insert the first protection disk on the spacer.
3. Place the transformer over the protection disk.
4. Place the second protection disk over the transformer.
5. Place the metal plate over the assembly.
6. Insert the fixing screw in the spacer passing through the metal plate.
Installation Guide
1. Installation Position of the Transformer for NPD
Figure 1-18. Transformer Installation
1
Upgrade Kit
NPD Tandem Detector
23