Thermo Scientific NPD Tandem Upgrade Kit Installation Manual page 24

For trace gc ultra
Table of Contents

Advertisement

Upgrade Kit
NPD Tandem Detector
7. Screw until the transformer assembly is blocked.
The transformer has three cables and four female connectors marked "Px".
Connect the female connectors to the corresponding male connectors marked "Jx"
on the power control module PCB (J14, J15) and on the mother board PCB (J35,
J36, J37) depending where the NPD board has been installed.
The result of the operation is shown in Figure 1-20 on page 25.
24
1. Transformer
3. Spacer
5. Fixing Screw
Figure 1-19. Transformer Assembly Identification Parts
2
2. Protection disks
4. Metal Plate
Cable Connector
P15
P4
P35 or 36
P36 or 37
1
4
5
3
Connect To...
J15
J4
J35 or 36
J36 or 37
Getting Started
Installation Guide