Quick Start Wizard via the handheld programmer

1. Quick Start
1.1. Quick Start Wiz
• Point the programmer at the display from a maximum distance of 300 mm (1 ft), then
press RIGHT arrow
• Press RIGHT arrow
• Press RIGHT arrow
move directly to the next item.
• To change a setting, scroll to the desired item or key in a new value.
• After modifying a value, press RIGHT arrow
to move to the next item.
Language
Selects the language to be used on the LCD and takes effect immediately.
Options
Material
Selects the appropriate echo processing algorithms for the material. See Position detect
(2.5.7.2.) (Page 104) for more detail.
Options
dK < 3.0
1)
Response Rate
Sets the reaction speed of the device to measurement changes in the target range.
Options
Use a setting just faster than the maximum filling or emptying rate (whichever is greater).
Units
Sensor measurement units.
Options
52
to activate PROGRAM mode and open menu level 1.
twice to navigate to menu item 1.1 and open parameter view.
to open Edit mode or DOWN arrow to accept default values and
English, Deutsch, Français, Español
LIQUID
LIQUID LOW DK
(low dielectric liquid – CLEF algorithm enabled)
1)
Response Rate (2.4.1.)
SLOW
MED
FAST
m, cm, mm, ft, in
to accept it and press DOWN arrow
Fill rate per Minute (2.4.2.)/Empty rate per
Minute (2.4.3.)
0.1 m/min (0.32 ft/min)
1.0 m/min (3.28 ft/min)
10.0 m/min (32.8 ft/min)
SITRANS LR200 with PROFIBUS PA