Quick Start Wizard Via The Handheld Programmer - Siemens sitrans LR200 Instruction Manual

2-wire, 6 ghz pulse radar level transmitter
Hide thumbs Also See for sitrans LR200:
Table of Contents

Advertisement

Quick Start Wizard via the handheld programmer

Notes:
• The Quick Start Wizard settings are inter-related and changes apply only after you
select Yes in Quick Start parameter Apply Changes (1.7).
• Do not use the Quick Start wizard to modify individual parameters: see instead
Parameter Reference
has been completed.)
• Each time the Quick Start Wizard is initiated, the start-up settings are factory
defaults. The Wizard will not recall previous user-defined settings.
1. Quick Start
a.
Point the programmer at the display (from a maximum distance of 300 mm [1 ft]),
then press RIGHT Arrow
b.
Press RIGHT Arrow
view.
c.
Press RIGHT Arrow
values and move directly to the next item.
d.
To change a setting, scroll to the desired item or key in a new value.
e.
After modifying a value, press RIGHT Arrow
Arrow
f.
Quick Start settings take effect only after you select Yes to Apply changes in step 1.7.
1.1. Material
Options
a)
dK < 3.0.
1.2. Response Rate
Sets the reaction speed of the device to measurement changes in the target range.
Options
Use a setting just faster than the maximum filling or emptying rate (whichever is
greater).
1.3. Sensor Units
Select the units for the Quick Start variables (high and low calibration point, and level,
distance, or space).
Options
Page 24
SITRANS LR200 (PROFIBUS PA) – INSTRUCTION MANUAL
on page 44. (Perform customization only after the Quick Start
to activate PROGRAM mode and open menu level 1.
twice to navigate to menu item 1.1 and open parameter
to open Edit mode or DOWN Arrow
to move to the next item.
LIQUID
LIQUID LOW DK (low dielectric liquid)
SLOW
0. 1 m/minute
MED
1.0 m/minute
FAST
10.0 m/minute
M, CM, MM, FT, IN
to accept default
to accept it and press DOWN
a)
7ML19985JR01