Automated Control - Thermo Scientific Dionex UltiMate 3000 Series Operating Instructions Manual

Hide thumbs Also See for Dionex UltiMate 3000 Series:
Table of Contents

Advertisement

UltiMate 3000 Series:
LPG-3400XRS Pump
3. Verify that the pump is connected to Chromeleon (the LED next to the Connect
button is green). If it is not, click Connect.
The control panel provides access to the operating parameters and commands required for
routine operation of the pump. Additional functions are available in the Commands dialog
box. To open the Commands box from the panel tabset, select Command on the Control
menu.

5.3.3 Automated Control

With automated control, you create a program file (PGM) for automated operation of the
pump. Programs can be created automatically with the help of a software wizard or
manually by editing an existing program. For details, see the Chromeleon Help.
To create a program with the Program Wizard
1. Open the Program Wizard. On the File menu, select New, and then select Program
File.
2. The wizard guides you through program creation. On each wizard page, make the
desired settings or accept the standard values. For additional information about a page,
click Help.
3. After you finish the wizard, Chromeleon automatically creates the corresponding
program.
4. To start the program, follow the steps on page 73.
To create a program manually
1. Open an existing program.
Select and double-click the program you want to open.
—or—
On the File menu, select Open. In the dialog box, select Program on the Object of Type
list and select the program.
Opens the
Commands View
Open the
Device Views
Fig. 34: Chromeleon program (here program shown in the Commands view)
Page 72
Operating Instructions

Advertisement

Table of Contents
loading

This manual is also suitable for:

Dionex ultimate lpg-3400xrs

Table of Contents