8 Technical Information

Operating
Serial dual-piston pump
principle
Compressibility
Fully automated, independent of the composition of the mobile phase
compensation
Flow rate range
Recommended
10 - 2,000 µL/min
Settable
1 - 2,000 µL/min
Flow rate
+/- 0.5 % or 1 µL/min, whichever is greater
accuracy
Flow rate
< 0.05 % RSD or < 0.01 min SD, whichever is greater
precision
Pressure range
2 - 125 MPa (18037 psi)
Pressure ripple
Typically < 1 % or < 0.2 MPa, whichever is greater
Gradient
Low-pressure gradient proportioning
formation
Proportioning
± 0.5 % (of full scale)
accuracy
Proportioning
< 0.15 % SD
precision
Number of
4
solvents
Gradient delay
Minimum of 200 µL (by default 200 µL; with optional dynamic mixer 235 µL or
275 µL, depending on the mixer)
volume
Solvent degassing 4 degassing channels
Max. flow rate 10 mL/min
Degassing capacity ~2 ppm residual gas at 1 mL/min
Delay volume per channel ~480 µL
Biocompatibility
No
Weight
15.8 kg
Dimensions
20 x 42 x 51 cm (h x w x d)
PC connection
All functions controllable via USB (USB 2.0) or RS-232
I/O
8-pin I/O terminal connector
Interfaces
Safety features
Leak sensor, active rear seal wash system, excess pressure monitoring
User input/display LCD indicating system parameters
Standby button
3 LEDs (Power, Connected, and Status) for status monitoring
GLP features
In Chromeleon: Support of automatic equipment qualification (AutoQ™ routines) and
System Wellness monitoring. All system parameters are logged in the Audit Trail.
Operating Instructions
UltiMate 3000 Series:
LPG-3400XRS Pump
Page 135