3.4.4.2 Connecting an OAS-3x00TXRS Autosampler
To connect the OAS-3x00TXRS autosampler to the pump, the following cables are required:
Cable
Autosampler interconnect cable —included in the autosampler shipment
For information about the connectors, see page 39.
Synchronization cable to connect the autosampler to the pump for pump stroke
synchronization
RS-232 cable to connect the autosampler to the RS-232 port on the Chromeleon data
system computer for inject response transmission
RS-232-USB Interface cable required in addition to the RS-232 cable if an RS-232 port
is not available on the computer
All cables are included in the accessories kit for the autosampler.
No.
Description
1
15-pin Sub-D connector, connects the appropriate port on the autosampler
2
25-pin Sub-D connector, connects the synchronization cable
3
2-pin connector, connects, for example, to the RS-232 synchronization cable to connect the
autosampler to the RS-232 port on the Chromeleon data system computer and/or to the Start In
pins of a mass spectrometer
For further information about the cables, refer to the Operating Instructions for the
autosampler.
1. Locate the autosampler interconnect cable (part no. 6043.0004) and the
synchronization cable (part no. 6043.0002). Both cables are shipped with the
autosampler.
2. Connect the 25-pin Sub-D connector of the interconnect cable to the 25-pin Sub-D
connector of the synchronization cable.
Operating Instructions
Fig. 18: OAS-3x00TXRS interconnect cable
UltiMate 3000 Series:
LPG-3400XRS Pump
Part No.
6043.0004
6043.0002
6043.0005
6073.2000
Page 39