Cleaning, Servicing And Despatching

Cleaning

Clean the device only when it is switched off and the electrodes have been discon-
nected. For this, first remove the power module from the device or remove the plug
(in case of the optional power pack) from the socket!
Do not use a dripping wet cloth for cleaning. Do not pour
any liquids on the device and / or do not dip it into any liquids!
Clean the PRIMEDIC™ DefiMonitor and all accessories such as the Wall Mount Rack
with any commercially available household cleaner.
Use a slightly damp and clean cloth for cleaning.
Use a standard disinfectant solution for disinfecting
(e. g. Gigasept FF)
Clean the paddles after each application. Soiling due to gel deposits can result in haz-
ards to the user and the patient. In any case, remove the gel residue from the pad-
dles.

Servicing

We recommend a regular inspection / servicing of PRIMEDIC™ DefiMonitor and all
accessories irrespective of the frequency of use by the user / service technician at
least once a year.
Check whether the casing, the cable, the SavePads and the accessories are not dam-
aged in any way.
Check List
Check the expiry date
– the SavePads,
– AkuPak and
replace the parts if required!
Check if
– the status display shows 'OK'
– the device can be switched on!
– the device automatically runs the self-test after it has been switched on!
– the compartment of the power supply unit is clean!
– the device is fully equipped!
– the internal battery is older than 10 years