Configuration parameters
October 2022
Figure B-8: Bottom Offset
A
B
C
A. Device Reference Point
B. Reference Height
C. Zero Level
D. Bottom Offset
B.6

Volume flow

B.6.1
Volume flow calculation method
Select the preferred volume flow calculation method. The are three options to choose
from:
Linearization table
Parshall flume
Khafagi-Venturi flume
Related information
Set up the volume flow measurement
B.6.2
Volume flow table
The volume flow table is used to convert the measured level into a volume flow rate. Up to
30 level-volume flow pairs can be entered. The level points must be entered in increasing
order.
88