Geometry - Emerson Rosemount 1208A Reference Manual

Level and flow transmitter
Hide thumbs Also See for Rosemount 1208A:
Table of Contents

Advertisement

Configuration parameters
October 2022
Related information
Analog saturation levels
B.5

Geometry

B.5.1
Advanced
Calibration offset
Difference between surface distance measured by transmitter and the same distance
measured by, for example, hand-dipping with a measurement tape. A positive Calibration
Offset value will increase the presented level value.
Figure B-6: Calibration Offset
A
B
A. Positive Calibration Offset value
B. Negative Calibration Offset value
C. Reported level
D. Actual level
Upper null zone
The Upper Null Zone defines how close to the device's reference point a level value is
accepted. You can extend this value to block out disturbing echoes close to the antenna,
for example from the tank nozzle.
Note
Make sure the 20 mA value is below the Upper Null Zone. Measurements are not
performed within the Upper Null Zone (UNZ).
86
Reference Manual
00809-0100-7062
C
D
Rosemount 1208A Level and Flow Transmitter