Test Procedure; Alternate Safety Test Procedures; Test Equipment - GE Responder 2000 Service Manual

Defibrilator/monitor
Hide thumbs Also See for Responder 2000:
Table of Contents

Advertisement

Test Procedure

Note: This procedure assumes the use of a Metron QA-90 Safety Analyzer. If an equivalent safety analyzer is used, refer
to the Operator's manual for detailed instructions.
1.
Press Responder 2000 Power button to turn on.
2.
Turn on the Safety Analyzer and wait for the self test to complete.
3.
If previously saved, select the stored Responder 2000 parameters.
4.
Press START on the Safety Analyzer to begin the tests.
5.
After test sequence is complete, verify TEST PASSED is displayed on the Safety Analyzer.

Alternate Safety Test Procedures

The following procedures are provided in case the recommended test procedures (above) cannot be performed.

Test Equipment

Number
Fluke Model Index 2 Sp02 simulator
Fluke Model ECG Simulator Defib analyzer or adequate
Metron QA-90 Safety Analyzer or adequate
MarqIII-Kit
Marq-3 EKG Simulator or adequate
2025269-003
RESPONDER 2000/PH ECD PACER TESTER
2025653-048 Revision B
Section 5: Performance Verification and Safety Testing
Equipment Description
Responder
2000
Qty
1
1
1
1
1
Page 75

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents