Repair Tools And Equipment; Troubleshooting Guide; General Troubleshooting - GE Responder 2000 Service Manual

Defibrilator/monitor
Hide thumbs Also See for Responder 2000:
Table of Contents

Advertisement

Repair Tools and Equipment

The following tools are needed to perform the procedures in this section.
TR15 Safety Torx
TR25 Safety Torx
Serial cable for software upgrades (with Windows 2000 or XP PC)
Software upgrade kit

Troubleshooting Guide

General Troubleshooting

Before performing any troubleshooting, check the following:
Ensure all external cables and connections are tight and undamaged.
If possible, ensure the unit is plugged in to a receptacle with appropriate voltage available (see SECTION 8:
Specifications and Safety for voltage requirements). If no receptacle is available, use a known good, charged
battery to perform any tests.
Visually inspect the unit for any obvious external damage, including cracks in the display, case, or connectors.
Ensure all connector pins and sockets are clean, free of debris, and intact.
Discuss the issues with the operator:
Have the operator explain and demonstrate the problem.
o
Ask about any previous repairs or problems.
o
Has the unit been stressed (including extreme heat/cold, submersion, falls, etc.)?
o