GE Responder 2000 Service Manual page 84

Defibrilator/monitor
Hide thumbs Also See for Responder 2000:
Table of Contents

Advertisement

Test Description
Patient leakage current, Mains on Applied Part (All SIP/SOPs Grounded): LEADS
MD connected to "LEADS" (all LEADS connected together), Type CF defibrillation proof limits
Caution: mains voltage on "LEADS"
Polarity: Normal
Neutral: Open
Ground: Normal (Closed)
Polarity: Reverse
Neutral: Open
Ground: Normal (Closed)
Patient leakage current, Mains on Applied Part (All SIP/SOPs Grounded): PADDLES
MD connected to "PADDLES" (all PADDLES connections tied together), Type CF defibrillation proof limits
Caution: mains voltage on "PADDLES"
Polarity: Normal
Neutral: Open
Ground: Normal (Closed)
Polarity: Reverse
Neutral: Open
Ground: Normal (Closed)
Patient leakage current, Mains on Applied Part (All SIP/SOPs Grounded): SpO
MD connected to "SpO
" (all SpO
2
Caution: mains voltage on "SpO
Polarity: Normal
Neutral: Open
Ground: Normal (Closed)
Polarity: Reverse
Neutral: Open
Ground: Normal (Closed)
2025653-048 Revision B
Section 5: Performance Verification and Safety Testing
S1: DPDT
S2: SPDT
Polarity
Neutral
a) Normal
a) Closed
b) Reverse
b) Open
Position
Position
a
b
a
X
X
X
X
connections tied together), Type BF defibrillation proof limits
2
"
2
X
X
Responder
S3: SPDT
S4: SPDT
Ground
Test point
a) Closed
a) Internal
b) Open
b) TP1
Position
Position
b
a
b
a
X
X
X
X
X
X
X
X
X
X
X
X
2
X
X
X
X
X
X
2000
Maximum Limits
(uA)
b
EN/IEC
UL
50
50
50
50
100
50
(60601-
2-4)
100
50
(60601-
2-4)
5,000
5,000
5,000
5,000
Page 84