Verify Operation

Before performing service on the Responder 2000, use the following procedures to configure and test the Responder
2000 to verify proper operation (or locate a fault condition):
Visual Inspection (page 59)
Operational Tests (starting on page 60)

Checkout Procedures

After service, the following operational and safety procedures must be performed to verify proper operation. All tests
indicated must be performed before the Responder 2000 can be returned to the customer. Depending on the type of
repair, the following procedures must be performed for each of the following cases:
External Repairs or Replacement
External repairs/replacements do not require opening the case. After repair or replacement, perform a Visual
Inspection (page 59).
Printer Door Replacement
The printer door may be replaced without opening the case. After the printer door is replaced (and the case was
not opened) the following tests must be performed:
1.
Visual Inspection (page 59)
2.
Printer Test (page 68)
Internal Repairs
If the case is opened, the following tests must be performed after the repair:
1.
Visual Inspection (page 59)
2.
Operational Check (starting on page 60)
3.
Safety Checks (starting on page 69)
2025653-048 Revision B
Section 5: Performance Verification and Safety Testing
Responder
2000
Page 52