GE Responder 2000 Service Manual page 78

Defibrilator/monitor
Hide thumbs Also See for Responder 2000:
Table of Contents

Advertisement

Earth leakage (AC Line)
Perform the following tests, with the applicable switch settings on the "Safety Tester" and record the results in the
applicable test form(s).
UUT power switch in the "ON" position.
Test Description
Earth leakage current measurements
TP1 – no connection
Polarity: Normal
Neutral: Normal (Closed)
Ground: Open
Polarity: Normal
Neutral: Open
Ground: Open
Polarity: Reverse
Neutral: Open
Ground: Open
Polarity: Reverse
Neutral: Normal (Closed)
Ground: Open
2025653-048 Revision B
Section 5: Performance Verification and Safety Testing
S1: DPDT
S2: SPDT
Polarity
Neutral
a) Normal
a) Closed
b) Reverse
b) Open
Position
Position
a
b
a
b
X
X
X
X
X
X
X
X
Responder
S3: SPDT
S4: SPDT
Ground
Test point
a) Closed
a) Internal
b) Open
b) TP1
Position
Position
a
b
a
b
X
X
X
X
X
X
X
X
2000
Maximum Limits
(uA)
EN/IEC
UL
500
300
1,000
1,000
1,000
1,000
500
300
Page 78