Safety Tests; Hipot Test; Required Equipment - GE Responder 2000 Service Manual

Defibrilator/monitor
Hide thumbs Also See for Responder 2000:
Table of Contents

Advertisement

Safety Tests

WARNING: Explosion or Shock Hazard.
Before performing any test, read and follow all safety instructions for the test equipment indicated.
WARNING: Explosion or Shock Hazard.
The following instructions are written for the test equipment indicated. Performing any test with equipment other than the
specific brand and model indicated may cause explosion, electrical shock, and equipment damage.
All safety tests are required whenever the case is opened, regardless of the type of maintenance or service performed.
Safety tests include:

Hipot Test

Leakage Current Test
Note: If the Hipot and Leakage current tests in this section cannot be performed for any reason, perform the Alternate
Safety Test Procedures starting on page 75.
Hipot Test
WARNING: Explosion Hazard.
Internal damage or improper assembly could cause an internal component to explode during Hipot testing. To minimize
risk
Ensure all Responder 2000 covers are in place and securely attached before testing
Use appropriate personal safety equipment, including a face shield and gloves
Ensure the area is clear before starting the test.
WARNING: Shock Hazard.
Do not touch the Hipot tester, the Responder 2000, or any leads during the test. Ensure all Hipot leads are electrically
isolated before starting the test. If the test is performed on an anti-static mat, ensure the leads do not contact the mat
during the test.

Required Equipment

UadTech Sentry 30 AC/DC/IR Hipot Tester
Test fixtures
ECG Connector Test Fixture. ECG connector with all leads and ground shield shorted together.
o
AC Plug Test Fixture. AC connector with line and neutral shorted together.
o
SpO
Connector Test Fixture. Connection for the ground shield.
o
2
ECG Paddle Test Fixture. Apex and Sternum pins shorted together.
o
Serial Port Test Fixture. Serial port connector with single lead connect to pin 5.
o
2025653-048 Revision B
Section 5: Performance Verification and Safety Testing
Responder
2000
Page 69

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents