J102
PDLHV_APEX
J104, J105, J106, J107 – Energy Storage Capacitor Connectors
Pin
Number
J104
HVCAP1_POS
J105
HVCAP1_NEG
J106
HVCAP2_POS
J107
HVCAP2_NEG
J126, J127 – Internal Test Load Resistor
Pin
Number
J126
TESTLOAD1
J127
TESTLOAD2

ECG PCBA

The ECG PCBA is a monitoring ECG subsystem. All operations are performed per serial communications with the main
CPU PCBA.
J412 – Interface to main CPU PCBA
Power and serial communications are carried through this connector.
Pin
Number
1
GND
2
SER_CPU2ECG
3
GND
4
SER_ECG2CPU
5
GND
2025653-048 Revision B
Patient Connection – Apex
Keystone 1287-ST
Name
Energy Storage Capacitor 1 – Pos
Energy Storage Capacitor 1 – Neg
Energy Storage Capacitor 2 – Pos
Energy Storage Capacitor 2 – Neg
Keystone 1287-ST
Name
Connects to a paddle cradle
Connects to a paddle cradle
SAMTEC BCS-104-L-D-PE-BE
Name
System Ground
Logic Level Serial Data Input
System Ground
Logic Level Serial Data Input
System Ground
Responder