Test Description
Polarity: Reverse
Neutral: Normal (Closed)
Ground: Normal (Closed)
Patient leakage current measurements: PADDLES
TP1 connected to "PADDLES" (all PADDLES connections tied together), Type CF defibrillation proof limits
Polarity: Normal
Neutral: Normal (Closed)
Ground: Normal (Closed)
Polarity: Normal
Neutral: Open
Ground: Normal (Closed)
Polarity: Normal
Neutral: Normal (Closed)
Ground: Open
Polarity: Reverse
Neutral: Normal (Closed)
Ground: Open
Polarity: Reverse
Neutral: Open
Ground: Normal (Closed)
Polarity: Reverse
Neutral: Normal (Closed)
Ground: Normal (Closed)
2025653-048 Revision B
Section 5: Performance Verification and Safety Testing
S1: DPDT
S2: SPDT
Polarity
Neutral
a) Normal
a) Closed
b) Reverse
b) Open
Position
Position
a
b
a
b
X
X
X
X
X
X
X
X
X
X
X
X
X
X
™
Responder
2000
S3: SPDT
S4: SPDT
Ground
Test point
a) Closed
a) Internal
b) Open
b) TP1
Position
Position
a
b
a
b
X
X
X
X
X
X
X
X
X
X
X
X
X
X
Maximum
Limits
(uA)
EN/IEC
UL
10
10
10
10
50
50
50
50
50
50
50
50
10
10
Page 81