Saving And Recalling Under Remote Operation; Using The Laser Compliance Voltage Adjustment - ILX Lightwave LDX-32420 User Manual

High power precision current source
Table of Contents

Advertisement

5
C H A P T E R
Releasing the knob for three seconds completes the RECALL operation. The
instrument is restarted and the front panel parameters are reconfigured to the new
parameters saved under that bin number.
The factory default values for front panel parameters may be restored by recalling
"bin" zero (0). See Table 2.1 for the default settings.

Saving and Recalling Under Remote Operation

For saving under remote operation, use the command *SAV x, where x is the
desired "bin" number (1-10). For a recall under remote operation, use the
command *RCL x, where x is the desired "bin" number (0 - 10).

Using the Laser Compliance Voltage Adjustment

For applications where you need to have critical protection of the laser, we
recommend the following. Using a resistor or other non-critical load, set the
LASER voltage limit as described in Chapter 2.
Then with the LASER output on, slowly increase the LASER current and watch
the LASER voltage measurement. Press the (DISPLAY) SELECT push button to
see the voltage. At about 0.25 V below the LASER voltage limit value, the
VOLTAGE LIMIT warning indicator will become lit on the display. Continue to
adjust the current upwards until the voltage limit is reached and the output is shut
off. Record the voltage at the value that the LASER output shuts off. It should be
within the specification of the LASER voltage limit accuracy. However, there is
typically some offset error. For example, if the LASER voltage limit is set to 4.0 V,
the actual voltage where it shuts off may be 3.9 V. This error term of 0.1 V should
be considered when setting the LASER compliance voltage limit.
The closer the LASER voltage limit shut off point is to the operating voltage of
your laser, the faster the circuit will work in the event of an open circuit. Some
experimentation may be necessary for optimum results.
68
LDX-32420
FU N C T I O N S A N D FE A T U R E S
Saving and Recalling Under Remote Operation

Advertisement

Table of Contents
loading

Table of Contents