ILX Lightwave LDX-32420 User Manual page 69

High power precision current source
Table of Contents

Advertisement

4
C H A P T E R
LASer:DISplay:PARAM
Parameters
Notes
Examples
LASer:DISplay:SET
LASer:DISplay:SET?
Notes
Examples
52
LDX-32420
C O M M A N D R E F E R E N C E
LDX-32420 Device-Dependent Command Reference
The LASer:DISplay:PARAM command enables the display to show the parameter
values.
None
This command has the same effect as pressing the (PARAMETER) SELECT switch. The
selected parameter will be displayed for three seconds. The actual LASER PARAM
display is turned off automatically when a display selection is enabled.
Each time the command is issued, the next LASER parameter will be selected, see
Chapter 2 for more information.
"Laser:Display:Param" - action: selects a parameter and displays its value.
The LASer:DISplay:SET command sets the display to show the setpoint of the selected
display mode.
The LASer:DISplay:SET? query returns the status of the setpoint display mode.
- where the response is 0 or 1.
LASer:DISplay:SET has the same effect as physically holding the (DISPLAY) SELECT
switch down continuously.
The setpoint display will not time out when REMOTE operation is used. (It will be
continuously displayed.)
"Las:Dis:Set" - action: enables the display for the setpoint of the selected mode: LDI,
MDI (I
) or MDP (P
PD
"LAS:DISPLAY:SET?" - response: 0, means the setpoint is not shown on the display.
)
PD