LDX-32420 Device-Dependent Command Reference
The LASer:CALMD command sets the laser's photodiode feedback responsivity, the
CAL PD parameter.
The LASer:CALMD? query returns this setting.
Responsivity:, in A/mW. The range is 0.01 to 100.00.
If the parameter is set to 0, the LDX-32420 will operate in a constant I
Constant P (MDP) mode is selected.
The parameter should be set to 0 for I
parameter is used to convert between I
A/mW.
If this value is 0, the front panel P
"LAS:CALMD 0.1" - action: sets the CAL PD parameter to 0.1. Therefore, a measured
PD current of 100A will equate to an optical power of 1000 mW (1W).
"LASER:CALMD?" - response: 0, means the LDX-32420 is set for operation in constant
I
mode (if P mode is also selected).
PD
The LASer:CAL:LDI command is used to enter the LASER current setpoint,
measurement, and limit calibration mode.
The LASer:CAL:LDI? query is used to determine that the LDX-32420 is ready for a value
to be entered during the calibration cycle of the LASer:CAL:LDI mode.
Since the limit circuit is the same for both high and low bandwidth modes, it is only
calibrated when low bandwidth mode is selected.
After LASer:CAL:LDI is issued, the LDX-32420 will allow calibration of the current
setpoint, measurement, and limit (if low bandwidth mode is selected). This procedure is
outlined in Chapter 6.
The calibration defaults to the selected bandwidth setting. Therefore, it is necessary to
select the desired bandwidth (I or I
the calibration for that bandwidth.
Calibration is performed at the current setpoint, wherever it is set. If the LASER output is
not on, E516 is displayed.
"Las:CAL:LdI" - action: the LDX-32420 enters calibration mode for LASER current.
LASer:CAL:LDI?" - response: 1, means the LDX-32420 is ready for the user to enter a
current value via the LASer:LDI command.
operation modes. Otherwise, the value of this
PD
and P
values. The units of this parameter are
PD
PD
will display "-.-" and no P
PD
) and turn the LASER output on before performing
HBW
F
P
RONT
ANEL
R
EMOTE
mode, when
PD
value can be calculated.
PD
F
P
RONT
ANEL
R
EMOTE

Advertisement

Table of Contents
loading

Table of Contents