For example, the user may enable the current source output (LAS:OUT 1),
enable an SRQ on Operation Complete (set *ESE 1 and *SRE 5), and have an
SRQ handling routine in the user's software, that reads a new measurement after
the output on state has been reached. This allows the use of the operation
complete features of the LDX-32420, without the need for program looping or
polling that can tie up the communication bus. Operation Complete on the
LDX-32420 is defined as:
1
2
Standard
Event Status Register
*ESR?
0 - Operation Complete
7
6
5
4
3
2
1
0
1 - Request Control
2 - Query Error
&
&
3 - Device-Dependendent Error
&
LOGICAL
&
&
4 - Execution Error
OR
&
5 - Command Error
&
&
6 - User Request
7 - Power On
7
6
5
4
3
2
1
0
Standard
Event Status Enable
Register
*ESE
*ESE?
SERVICE
REQUEST
GENERATION
LOGICAL
OR
QUEUE
NOT-EMPTY
{
ERROR
QUEUE
Note: If the GPIB times out while waiting for a response, either set the GPIB time-our longer or
34
LDX-32420
R E M O T E O P E R A T I O N
Status Reporting
No operations to the LASER current source hardware are pending.
No EEPROM (non-volatile) memory write cycles are in progress.
QUEUE
NOT-EMPTY
{
OUTPUT
QUEUE
Status Byte Register
*STB?
{
0 – N/A
7
6
5
4
3
2
1
0
1 - N/A
&
2 – LASER Event
&
3 – LASER Cond.
&
&
4 - Message Available
&
&
5 - Event Status Summary
&
6 - Request Service/Master Status Summary
7 - Error Available
{
7
5
4
3
2
1
0
Service Request Enable
Register
*SRE
*SRE?

Figure 3.2 LDX-32420 Status Reporting Schematic Diagram

use SRQ generated interrupts in your program. See your GPIB manual for
time-out configuration or SRQ programming setup.
The *OPC, *OPC? and *WAI commands should not be used inside a calibration routine.
L ASER Condi t i on St at us Re gi s t er
L ASer : COND?
15
14
13
12
11
10
9
8
7 6 5 4 3 2 1 0
&
&
&
&
&
&
&
&
LOGI CAL
&
&
OR
15 14 13 12 11 10 9 8
7
6
L A SER Condi t i on St at us Enabl e
L A S : ENABl e : COND
L AS : ENA Bl e :
15 14 13 12 11 10 9 8
&
&
LOGI CAL
OR
15
14
0 – Current Limit
1 – Voltage Limit
2 – N/A
3 – Power Limit
4 – Interlock
5 – N/A
6 – Comp. Voltage Limit
7 – Open Circuit
8 – Output Shorted
&
9 – Out of Tolerance
&
&
10 – Output On
&
&
11 – Ready for Calibration
&
12 – N/A
13 – N/A
5
4
3
2
1
0
14 – N/A
15 – N/A
L A SER Ev ent St at us Re gi s t er
L ASer : EV Ent ?
7
6
5
4
3
2
1
0
&
&
&
&
&
&
&
&
&
&
&
&
&
&
13
12
11
10
9
8
7 6 5 4 3 2 1 0
L ASER Ev e nt St at us Enabl e
L AS : ENABl e : EVEnt
L A S : ENA Bl e :
0 – Current Limit
1 – Voltage Limit
2 – N/A
3 – Power Limit
4 – Interlock
5 – N/A
6 – Comp. Voltage Limit
7 – Open Circuit
8 – Output Shorted
9 – Tolerance Change
10 – Output On/Off change
11 – New Measurement Ready
12 – N/A
13 – N/A
14 – N/A
15 – N/A