ILX Lightwave LDX-32420 User Manual page 65

High power precision current source
Table of Contents

Advertisement

4
C H A P T E R
LASer:COND?
Notes
Examples
48
LDX-32420
C O M M A N D R E F E R E N C E
LDX-32420 Device-Dependent Command Reference
The LASer:COND? query returns the value of the status condition register of the LASER
operations.
-where the response is the sum of the following:
1
LASER Current Limit
2
LASER Voltage Limit
4
N/A
8
Power limit
16
Interlock disabled
32
N/A
64
Compliance Voltage Limit
128
Open circuit
The LASER conditions that are reported to the status byte are set via the
LASer:ENABle:COND command.
The Open circuit condition is only present while the LASER output is on, and when the
hardware detects this condition, it will turn the LASER output off. Therefore, the Open
Circuit condition is fleeting and may be missed via the LAS:COND? query. Therefore, the
user should test for the Open Circuit Event via the LAS:EVEnt? query.
The LASER condition status is constantly changing, while the event status is only
cleared when the event status is read or the *CLS command is issued.
"LAS:COND?" - response: 513, means that the LASER limit current and out of tolerance
LASER conditions currently exist.
"Radix Hex; Laser:Cond?" - response: #H108, means that the LASER Output shorted
and Power limit conditions currently exist.
256
Output is shorted
512
Output is outside tolerance limit
1024
Output on/off state
2048
Ready for calibration data state
4096
N/A
8192
N/A
16384
N/A
32768
N/A
F
P
RONT
ANEL
R
EMOTE