LASer:CAL:LDV
LASer:CAL:LDV?
Notes
Examples
LASer:CAL:MDI
LASer:CAL:MDI?
Notes
Examples
01_10
LDX-32420 Device-Dependent Command Reference
The LASer:CAL:LDV command is used to enter the LASER voltage measurement
calibration mode.
The LASer:CAL:LDV? query is used to determine that the LDX-32420 is ready for a
value to be entered during the calibration cycle of the LASer:CAL:LDV mode.
After LASer:CAL:LDV is issued, the LDX-32420 will allow calibration of the laser voltage
measurement. This procedure is outlined in Chapter 6.
"Las:CAL:Ldv" - action: the LDX-32420 enters calibration mode for LASER voltage.
"LASer:CAL:LDV?" - response: 1, means the LDX-32420 is ready for the user to enter a
voltage value via the LASer:LDV command.
The LASer:CAL:MDI command is used to enter the LASER photodiode current
calibration mode.
The LASer:CAL:MDI? query is used to determine that the LDX-32420 is ready for a value
to be entered during the calibration cycle of the LASer:CAL:MDI mode.
After LASer:CAL:MDI is issued, the LDX-32420 will automatically enter the LASER
photodiode current calibration mode. When the LDX-32420 is ready, the user should
enter the measured value. This procedure is outlined in Chapter 6.
If the LASER output is not ON or if the P mode is not selected, error code E-516 will be
generated.
"Las:CAL:MDI" - action: the LDX-32420 enters the LASER photodiode current
calibration mode.
LAS:CAL:MDI?" - response: 1, means the LDX-32420 is ready for the user to enter a
photodiode current value via the LASer:MDI command