Output Off Register
The Output Off Enable Register allows you to determine that conditions and
events in the LDX-32420 High Power Precision Current Source can cause the
current source output to be turned off. This register is configured in a manner
similar to the status reporting registers. However, the register output is not
reported in the Status Byte Register. Rather, it sets the hardware that controls the
output switching. The events and conditions that may be set to cause the current
source output to be turned off are shown in Figure 3.3.
The default (factory) settings for this register are shown in Table 3.3. These
settings are not affected by the *PSC (Power-On Status Clear) command.
Table 3.3 LDX-32420 Series Default Settings for Output Off Registers
0 - disabled
1 - disabled
2 - N/A
3 - enabled
4 - N/A
5 - N/A
6 - N/A
7 - N/A
01_10
8 - N/A
9 - disabled
10 - N/A
11 - N/A
12 - N/A
13 - N/A
14 - N/A
15 - N/A
R E M O T E O P E R A T I O N
Status Reporting
3
C H A P T E R
35