MAINTENANCE

16
WATER TRAP, PARTICLE & WATER STOP FILTER
16.1
Some boilers produce high water vapour volume which can affect your
analyser.
Your analyser has a water trap & particle filter to stop water vapour & dust.
Your analyser also has a water stop filter with hydrophobic technology
located inside the particle filter connected to a red filter carrier.
Drain water trap when you see ater collecting in the water trap - Remove red
plug, empty water then replace red plug.
Replace water stop & particle filters when they are wet or dirty or your
analyser displays LOW FLOW.
To replace:
Disconnect water trap from your analyser
1.
Remove red filter carrier with particle & water stop filter from water trap
2.
Ensure water trap is completely dry
3.
Attach new water stop & particle filter to red carrier
4.
Insert red carrier in water trap and reconnect water trap to your analyser.
Replacement parts:
Water trap: SM50515
Particle filter: PF400/5 (pack of 5 particle filters)
Water Stop filter: SM50933 (Needs a pack of 5)
Red filter carrier: CM50302
FLUE GAS & TEMPERATURE PROBE
16.2
Check your flue gas & temperature probe for tubing leaky crack.
Check analyser connectors are not bent or cracked.
Check flue gas temperature probe is not bent or out of shape,
WARNING
Never cool your probe in water or use probe shaft as a lever.
40 KANE458s MANUAL