CONTINUED MENU CHOICES
NOX CALC
LOGS
EFF
UTIL
CODE
(If NO sensor fitted)
Set REFERENCE NOx to percentage
required or defined by local regulations.
It is typically assumed boiler emissions of
NOx = NO + 5% NO2
If required, you can calculate NOx from 3
options - SUM, NO or NO2.
NOX CALC
SUM adds NO & NO2 if both sensors are
fitted.
NO calculates NOx from NO, if fitted, where
NOx = NO x 1.1
NO2 calculates NOx from NO2 only, if fitted,
where NOx = NO2 x 2.05.
Use
LOGS
View current memory usage & stored logs
Select EFF to set efficiency scale by
pressing ENTER.
Select NETT or GROSS using
EFF
press ENTER to confirm
Considering calculation automatically
selected based on selected fuel
Select UTIL to access utility menu - Press
ENTER then
from:
INFO - Shows Analyser firmware, wireless
firmware & date next calibration due
LEAK - Performs system leak test following
UTIL
onscreen instructions
B'LIGHT - Selects time backlight is on from
15 to 300 seconds using
to confirm
NOTE: Extending backlight time reduces
analyser battery life
Password protected for authorised service
CODE
agents only - Default to 000000
or
to select then press ENTER
or
& ENTER to select
KANE458s MANUAL 19
&
, then
or
& ENTER