10. MAINTENANCE

EMPTYING AND CLEANING THE IN-LINE WATER TRAP

The in-line water trap should be checked and emptied on a regular basis.
Water vapour will condense and gather in the probe line. This may move
suddenly to the trap when the probe is moved. Care should be taken at
all times.
Emptying of the water trap is detailed below :-
Carefully remove the end cap from the in-line housing. Dispose of the
condensate in a suitable drain, care must be taken as it could be acidic.
If condensate spills onto the skin or clothing, clean off immediately using
fresh water, seek medical advice if problems occur.

CHANGING THE PARTICLE FILTER

This is a very important part of the analyser and should be changed
regularly. It prevents dust and dirty particles entering the pump and
sensors and hence causing damage. The filter MUST be changed when
it is discoloured.
Remove the end cap from the in-line filter housing. Carefully remove the
paper filter element and dispose of it. Clean the inside of the filter
housing with a suitable soft cloth. Insert a new filter element onto the
spigot in the filter housing and carefully replace the end cap.
KANE905 Operators Manual
Page 33