LET-BY / TIGHTNESS TESTING & EDITING TIMES
12.3
Rotate dial to TIGHTNESS then press ENTER to zero pressure sensor.
Using black connectors, connect your manometer hose from appliance test
point to analyser P1 input.
Display shows "LET BY?" - use
If YES selected, set let-by pressure then press ENTER to start let-by
test:
If let-by test fails rotate dial to another position to stop test.
If let-by test passes, adjust gas pressure for tightness test & press ENTER
to start stabilisation test - display shows:
When complete press ENTER to start tightness test:
32 KANE458s MANUAL
or
& ENTER to select YES or NO.
Let-by test automatically stored in
memory
Pressure at start of let-by test
Real time reading
Let-by default time: 60 seconds - can be
changed via "Menu"
Real time pressure reading
Stabilisation default time: 60 seconds