CONTENTS Page No. KANE458s OVERVIEW & KANE LINK ANALYSER AT A GLANCE 7-10 ANALYSER LAYOUT KEYPAD & ROTARY DIAL BACK OF ANALYSER & PROBE GENERAL SAFETY FLUE GASES PROTECTION AGAINST ELECTRIC SHOCK BATTERIES BATTERY TYPE REPLACING BATTERIES UPDATING TIME AND DATE...
Page 3
SELECT LOG TYPE SELECT LOG STORED SELECT LOG OPTIONS MEASURING FLUE GASES HIGH CO LEVELS VIEWING FLUE GASES USING ROTARY DIAL 10.1 AUX SCREEN 10.2 02 / EFF SCREEN CO / CO2 SCREEN 10.3 10.4 IAQ SCREEN KANE458s MANUAL 3...
Page 4
KANE79 PERSONAL & ROOM CO MONITOR 15.3 TRANSFER DATA TO SMARTPHONE OR TABLET 15.4 MAINTENANCE WATER TRAP, PARTICLE FILTER & WATER STOP 16.1 FILTER FLUE GAS & TEMPERATURE PROBE 16.2 BATTERY CHARGER & BATTERIES 16.3 SYSTEM LEAK TEST 16.4 KANE458s MANUAL 4...
Page 5
COLD WEATHER PRECAUTIONS WHERE TO SEND YOUR ANALYSER SPECIFICATION 43-44 EU DECLARATION OF CONFORMITY KANE CARE 46-49 KANE CARE 5 YEAR SERVICE PLAN COMBUSTION PROBE OPTIONS ASSESSORIES & CONSUMABLES ANALYSER RECYCLING THANK YOU KANE458s MANUAL 5...
Large 6 line display showing data & test results - Display bottom line also highlights analyser status. Test reports to optional infrared printer or wirelessly to KANE LIVE App. 45 log memory for any combination of Combustion, AUX, Temperature &...
Water trap drain with red plug Pressure port P1 Red connector Connect to Flue Gas Inlet Flue Gas Probe Temperature Plug - Connect to Temperature T1 Narrow (+ve) pin must be on right side Flue probe hose 8 KANE458s MANUAL...
PRINT LOG - Short press to transfer a log - Analyser offers a SEND destination choice Keypad Rotary Dial SYMBOL DESCRIPTION Add, Manage or Delete KANE LINK devices & App Settings MENU Manage Analyser Settings STATUS View Analyser Status Personalise Analyser Display O2 / EFF Appliance Efficiency Test...
ANALYSER BACK & PROBE Flue Magnets x 2 Temperature Probe Flue Gas Probe exhaust Depth stop cone Water trap drain with red plug Thermocouple wire Flue Gas Probe hose 10 KANE458s MANUAL...
Altitude to 2000m • Ambient temperature 0°C -40°C • Maximum relative humidity 80% for temperatures up to 31°C • decreasing linearly to 50%RH at 40°C Mains supply fluctuations not to exceed 10% of the nominal voltage • KANE458s MANUAL 11...
USER INTERFACE Your analyser displays 5 lines of tests & a status bar. The display backlight power on with each button press then powers off after 10 seconds. Navigate through options and menu choices via & ENTER. KANE458s MANUAL 13...
Current atmospheric pressure (mbar) No. of days until next calibration due Status bar STATUS BAR Status bar shows analyser status & offers options based on your settings Navigate through status bar options via when status bar on display. Status bar 14 KANE458s MANUAL...
CALIBRATION DUE Display alternates WARNING MESSAGE between calibration due & current time AIR PURGE Display alternates between “SNOOZE” air purge delay symbol & current time TIMER LOW BATTERY Display alternates WARNING between BAT symbol & MESSAGE current time KANE458s MANUAL 15...
STATUS BAR ICON ZONE Icons give sample status information: PUMP Pump status On or Off Analyser has logged / LOGGED DATA stored data SENDING DATA Data being sent to App or Printer BATTERY Battery status indication 16 KANE458s MANUAL...
& ENTER Menu option to edit As you navigate up or down items move up or down screen returning to the beginning. To exit MENU rotate dial to any position but note unsaved changes will be lost. KANE458s MANUAL 17...
ENTER HEADER HEADER To set HEADER via KANE LIVE app, select CONTROLS on app, select ANALYSER HEADERS, select LINE to edit Select SAVE to set header Select IR PRINTER to set infrared printer -...
Page 19
B’LIGHT - Selects time backlight is on from 15 to 300 seconds using & ENTER to confirm NOTE: Extending backlight time reduces analyser battery life Password protected for authorised service CODE CODE agents only - Default to 000000 KANE458s MANUAL 19...
SEND, PRINT OR STORE A TEST Short press SEND to send a test to your optional KANE-IRP3 printer or KANE LIVE App. Long press SEND for 2 seconds to store a test, called log. ANALYSER MEMORY Your analyser has a shared memory system to store up to 45 tests, called logs.
View Pressure & Temperature Logs Return to previous menu SELECT LOG STORED After you select report type first log is displayed: LOG Number of that type Time & date of LOG Report readings specific to log type Navigation menu options KANE458s MANUAL 21...
Do not exceed analyser internal temperature operating range • Do not put analyser on hot surfaces • Do not exceed water trap levels • Do not let analyser particle filter become dirty and blocked • Check readings are stable and within expected range. KANE458s MANUAL 23...
EDIT appears on status bar then long press and hold ENTER until cursor flashes and line number appears in status bar. to select option to appear on line then press ENTER. Repeat for other lines as required. 24 KANE458s MANUAL...
Excess air reading Pressure reading Status bar IAQ SCREEN 10.4 Rotate dial to IAQ to view Indoor air quality - CO only or CO & ambient CO2: CO ONLY Ambient CO reading Maximum recorded ambient CO reading KANE458s MANUAL 25...
If T2 is not connected, analyser internal temperature is used to calculate net temperature. Use T1 connection for flow temperature sensor Use T2 connection for return temperature sensor Real time temperature difference Real time pressure reading Status bar 26 KANE458s MANUAL...
11.3 LARGE BORE TUBING ISSUES If using large bore tubing when performing pressure tests: Push orange tube over rim of spigot to ensure gas tight seal: Failure may not produce gas tight seal KANE458s MANUAL 27...
TEST 4 - MEASURE FLOW & RETURN TEMPERATURES FROM APPLIANCE COMMISIONING tests are automatically stored in memory with a log number. Send COMMISIONING test log to your optional KANE-IRP3 printer by pressing ENTER or wirelessly to your KANE LIVE App using & ENTER 28 KANE458s MANUAL...
ALARM - 30ppm TYPE A WATER 5 minute test with results LIMIT = 10ppm HEATER stored every minute ALARM - 30ppm TYPE A SPACE 30 minute test with results LIMIT = 10ppm HEATER stored every minute ALARM - 30ppm KANE458s MANUAL 29...
Page 30
CO limit in ppm CO alarm limit in ppm CO & AMBIENT CO2 Test 1 of 15 Test 30 = maximum 30 tests in series Test interval time Ambient CO2 reading Ambient CO reading CO limit in ppm 30 KANE458s MANUAL...
Page 31
Otherwise it automatically stops after the pre-set time. ROOM tests are automatically stored in memory with a log number. Send ROOM tests to your optional KANE-IRP3 printer by pressing ENTER or wirelessly to your KANE LIVE App using & ENTER...
If let-by test passes, adjust gas pressure for tightness test & press ENTER to start stabilisation test - display shows: Real time pressure reading Stabilisation default time: 60 seconds When complete press ENTER to start tightness test: 32 KANE458s MANUAL...
RUN appears on status bar then press ENTER to start. TIGHTNESS tests are automatically stored in memory with a log number. Send TIGHTNESS test log to your optional KANE-IRP3 printer by pressing ENTER or wirelessly to your KANE LIVE App using &...
PRINTING TO INFRARED PRINTER To print test result or logs use optional KANE infrared printer. Power on printer and place printer infrared receiver in line with analyser emitter see page 8. Allow 15cm gap between analyser & printer.
SAMPLE SIMULTANEOUS ROOM CO TESTS WITH 14.1 KANE LINK CO MONITORS Multi room printouts are of individual reports for each room: Room 1 Room 2 Room 3 Room 4 36 KANE458s MANUAL...
Analyser or wirelessly transfer test results to your smartphone or tablet. Rotate dial to KANE LINK to manage how your analyser communicates with wireless devices or your smartphone or tablet. To ADD, REMOVE and check STATUS of any KANE LINK device select using & ENTER 15.1...
Belt clip Serial Number If shorter, enter 0’s to make up to 10 - e.g: in this example enter J12345678 as 0012345678. Other KANE LINK devices can be paired - Contact KANE for more details 38 KANE458s MANUAL...
Rotate dial of your KANE LIVE analyser to KANE LINK then select and enter APP mode using & ENTER. Open KANE LIVE then tap CONNECT to find your analyser - Select from device list then, if asked, tap PAIR to connect. KANE458s MANUAL 39...
Check your flue gas & temperature probe for tubing leaky crack. Check analyser connectors are not bent or cracked. Check flue gas temperature probe is not bent or out of shape, WARNING Never cool your probe in water or use probe shaft as a lever. 40 KANE458s MANUAL...
To perform, rotate dial to MENU Select UTIL pressing & ENTER Select LEAK pressing & ENTER Follow analyser instructions to connect and block probe Select NEXT pressing ENTER Your analyser now performs test then indicate pass or fail KANE458s MANUAL 41...
‘ON’ sampling fresh air for a few hours - Connect charger to avoid draining batteries. If you still experience problems please contact KANE Customer Service. WHERE TO SEND YOUR ANALYSER - See...
±0.2% of reading Pre-programmed Fuels UK USA & France Natural Gas, Propane, Butane, LPG, Light Oil, Wood Pellets Battery Life >8 hours (continuous with pump on) The KANE458s is independently test and certified to EN50379, Certification Parts 1-3 KANE458s MANUAL 43...
Page 44
SPECIFICATION CONTINUED Operating Conditions Temperatures 0 - 45°C Humidity 15 to 90% RH, (non-condensing) Power Supply Rechargeable batteries, USB Charging Physical Characteristics Weight Approx. 0.625g Dimensions 216mm x 105mm x 45mm 44 KANE458s MANUAL...
(EMC) 2014/35 Low Voltage Directive (LVD) The following harmonised standards and technical specifications have been applied: Certification The KANE458s is independently tested and certified to EN 50379, Parts 1 & 3 EN50270:2015 SAFETY EN61010-1:2010 RoHS (UK & EU)
Welcome to KANE CARE KANE CARE is our award winning promise to never let you down or your money back All gas analysers must be serviced and recertified every year - In some markets this is a legal requirement KANE CARE is all inclusive and applies to any KANE analyser booked online via www.kane.co.uk &...
Page 48
Select FREEPOST for tracked carriage - UK & Ireland mainland only Your analyser will be despatched the same day we receive it… OR YOUR MONEY BACK* *Excludes KANE-EGA, AUTO600, ‘9 series’ analyser & UKAS certificates - Ireland 2 day turnaround 48 KANE458s MANUAL...
Page 49
International All Inclusive Price KANE CARE Service & Recertification email: sales@kane.co.uk with analyser name & serial number Pay our all inclusive KANE CARE quote Send your analyser & accessories We will return free of charge* after KANE CARE Service & Recertification with 1 year warranty if under 10 years old *Excludes local tax &...
Purchase our 5 Year Service Plan with your KANE258, 358 or 458s standard configuration, covering years 1-4 service & recertification, saving money & giving 5 years of cover, with all the benefits of KANE All inclusive no quibble repairs Theft protection cover - if stolen we’ll replace your KANE...
Before use, please register on our website www.kane.co.uk Scan the QR code to go directly to Register your product online. Kane International Ltd Kane House, 11 Bessemer Road Welwyn Garden City Hertfordshire AL7 1GF, UK email: sales@kane.co.uk telephone: 0800 059 0800...