12.2
ROOM CO & CO2 TESTS
Rotate dial to ROOM.
Select ROOM CO using
for up to 30 minutes.
If fitted, ambient CO2 levels are also recorded.
Use
or
& ENTER to select test types from these options:
ROOM CO & CO2 TEST TYPES
TEST TYPE
GENERAL
SWEEP TEST
MIQRATION TEST
TYPE C SEALED
APPLIANCE
TYPE B BOILER
OPEN FLUE
TYPE A COOKER
TYPE A WATER
HEATER
TYPE A SPACE
HEATER
or
& ENTER to record room CO & CO2 levels
DURATION
15 minute test with results
stored every minute
2 minute test with results
stored every minute
15 minute test with results
stored every minute
15 minute test with results
stored every minute
15 minute test with results
stored every minute
30 minute test with results
stored every minute
5 minute test with results
stored every minute
30 minute test with results
stored every minute
LIMITS / ALARM
LEVELS
LIMIT = 10ppm
ALARM - 30ppm
LIMIT = 10ppm
ALARM - 30ppm
LIMIT = 10ppm
ALARM - 30ppm
LIMIT = 10ppm
ALARM - 30ppm
LIMIT = 10ppm
ALARM - 30ppm
LIMIT = 10ppm
ALARM - 30ppm
LIMIT = 10ppm
ALARM - 30ppm
LIMIT = 10ppm
ALARM - 30ppm
KANE458s MANUAL 29