GENERAL OPERATION The nominal pressure range of the meter is indicated on the label fixed to the back of the meter. When the meter is first switched on using , the top line of the display scrolls through the model version number, battery status, pressure units, date and time. The bottom line of the display shows a countdown number that starts at 10 and reduces to zero.
Page 3
UP / DOWN Press either to change the bottom line of the display. ENTER The buttons with arrows change settings such as date, time, and other menu items when MENU mode has been selected. Press to cancel OFF. button is also a shortcut to the let-by / stabilization / tightness test. (Only applies to KANE3500-1/UK and KANE3500-2/UK) BOTTOM LINE OPTIONS buttons change the bottom line of the display.
BEFORE USING THE METER FOR THE FIRST TIME: Remove the cover and fit a new battery in the battery compartment. Take great care to ensure that the battery is fitted with the correct battery polarity. Replace the battery cover. Always check that the meter is working correctly after replacing the battery.
USING THE MENU Press then navigate through the menu by using following keys… = Scroll = Scroll up = Enter down MENU SUB-MENU OPTIONS / COMMENTS Use the / and keys to view a previously logged test. STORE VIEW Other tests can then be viewed using the /...
Page 6
MEASURING Make sure you do not exceed the meter’s operating specifications. Do not exceed the meter’s internal temperature operating range Do not put the meter on a hot surface Switch the meter on with no hose connected to ensure that the initial ‘zero’ is accurately set.
Page 7
EXAMPLE PRINTOUTS The standard printout is: The MAX and MIN readings are those observed since last switch on NOTE: Printouts of stored readings will also include the TEST NO. below the header.
Page 8
LET-BY and TIGHTNESS TESTING (only applies to KANE3500-1/UK and KANE3500-2/UK) Switch the meter ON with no hose connected and wait for the 10 second countdown to finish. STEP DISPLAY COMMENTS / ACTIONS 0.00 Press to enter the let-by/stabilisation/tightness test sequence. This test sequence can also be entered via menu/pressure/report/test mbar LET BY?
Page 9
Let-By Test and Tightness Test Printout NOTE: Printouts of stored readings will also include the LOG NO. Below the header...
6. METER ANNUAL RECALIBRATION AND SERVICE The meter should be re-calibrated and serviced annually by a Kane approved service centre. Local regulations may require more frequent re-calibration. Calibration is performed in firmware and there are no user accessible adjustments or user serviceable parts.
Pressure Measurement WARNING - Maximum over-range without damage to sensor is 5 times nominal range Model: KANE3500-1 Range Resolution Accuracy + 20 Pa 0.1 Pa +0.5 Pa + 100 Pa 0.1 Pa + 3 Pa + 2000 Pa 1 Pa + 3% of reading + 80 hPa 0.01 hPa...
Page 14
Model: KANE3500-30 Range Resolution Accuracy + 1076 hPA 0.1 hPa + 10 hPa + 2070.0 hPA 1 hPa + 3% of reading Dimensions Weight 310 grams with battery Handset 160mm x 80mm x 40mm (180mm incl. spigots) Ambient +0C to +45C Operating 10% to 90% RH non- Range...
If not acceptable, adjust the meter’s position to minimize interference or switch off, if possible, the offending equipment during your test. At the time of writing this manual (June 2014) Kane International Ltd are not aware of any field based situation where such interference has occurred and this...
Page 16
ELECTROMAGNETIC COMPATIBILITY This product has been tested for compliance with the following generic standards: EN 61000-6-3:2011 EN 61000-6-1:2007 and is certified to be compliant Specification EC/EMC/KI/KANE3500 details the specific test configuration, performance and conditions of use. Please Note: Batteries used in this instrument should be disposed of in accordance with current legislation and local guidelines.
APPENDIX 1 – Main Parameters : UNITS : User selectable from : inH2O mbar mmH2O NOTE! Not all units available on all models. DATE : Date shown as day, month and year. The order can be changed using the menu function. Date is recorded when each test is printed.
Page 18
Once the meter has been securely packed then your package is ready for shipment back to Kane. If you do not have an account with a courier company you can take your package to your local Post Office. It is advisable to send the package by Special Delivery so that it is insured and traceable while in transit.
Page 19
Service Returns (Simply cut out and attach to your package) Northern Service Centre Kane International Ltd Gibfield Park Avenue Atherton Manchester M46 0SY Southern Service Centre Kane International Ltd Kane House, Swallowfield Welwyn Garden City Hertfordshire AL7 1JG Northern Service Centre...
Product Registration Please complete, detach and return to: Kane International Ltd, Kane House, Swallowfield, Welwyn Garden City, Hertfordshire, AL7 1JG Your Details Name: Job Title: Company Name: Company Address 1: Address 2: Town/City: County: Postcode: Country: Phone Number: Fax Number:...
Page 22
Why did you buy a Kane Product? Dealer Recommendation Other Recommendation Value for Money Our Fixed Price Servicing Programme Kane Previous Owner Not your Decision Other: What brand was your previous analyser?
Page 24
Thank you for buying this pressure meter. Before use, please register on our website www.kane.co.uk or complete, detach and return the Product Registration page.