For hazardous substances centrifuge containers with special screw caps must strictly be used. In addition to the
screw cap centrifuge containers, for materials in hazard category 3 and 4 a biosafety system must be used (see
the World Health Organisation's "Laboratory Biosafety Manual").
Without the use of a biosafety system the centrifuge is not microbiologically sealed in the sense of the
EN / IEC 610101-2-020 standard.
No biosafety systems are available for this centrifuge.
 The centrifuge must not be operated with highly corrosive substances which could impair the mechanical integrity
of rotors, hangers and accessories.
 Repairs must only be carried out by personnel authorised to do so by the manufacturer.
 Only original spare parts and original accessories licensed by the Andreas Hettich GmbH & Co. KG company are
allowed to be utilised.
 The following safety regulations apply:
EN / IEC 61010-1 EN / IEC 61010-2-020 as well as their national deviations.
 The safe operation and reliability of the centrifuge can only be guaranteed if:
 the centrifuge is operated in accordance with the operating instructions,
 the electrical installation on the site where the centrifuge is installed conforms to the demands of EN / IEC
stipulations.
 Meeting the country-specific requirements concerning occupational safety with regard to the use of laboratory
centrifuges at the workplaces provided for this purpose by the user is the responsibility of the user.
7

Transport and storage

7.1

Transport

Before transporting the device, the transport securing device must be installed.
When the device and accessories are transported, the following ambient conditions must be complied with:

Ambient temperature: -20°C to +60°C

Relative humidity: 10% to 80%, non-condensing
7.2

Storage

The device and the accessories may only be stored in closed and dry rooms.
When the device and accessories are stored, the following ambient conditions must be complied with:

Ambient temperature: -20°C to +60°C

Relative humidity: 10% to 80%, non-condensing
8

Scope of delivery

The following accessories are delivered with the centrifuge:
1 connecting cable
2 fuses
1 hex head wrench
1 operating manual
1 information sheet, transport securing device
1 information sheet, emergency unlocking
The rotor(s) and associated accessories are included in the delivery in the quantity ordered.
AB1101EN_SA