 Check whether the displayed maximum number of run
cycles matches the specifications on the
rotor/hangers.
If not, the maximum permissible number of run cycles
specified on the rotor/hangers must be set.
For rotors and hangers whose period of use is
not limited to a maximum number of run cycles
(centrifugation runs), zero must be set.
 Set the correct value with the keys.
 Press the key.
START
PULSE
The maximum permissible number of run
cycles is saved.
20

Settings and queries

PROG
RPM
t/min:s
Query:
 System information
(section 20.1)
24/45
Example:
PROG
PROG
PROG
PROG
RPM
t/min:s
PROG
RPM
t/min:s
Setting:
 Acoustical signal
(section 20.2)
 Automatic unlocking of
the lid after the
centrifugation run
(section 20.4)
 Automatic switch-off of
the background lighting
(section 20.5)
 Optical signal after
ending the
centrifugation run
(section 20.3)
 Displayed
centrifugation
parameters after
switching on the device
(section 20.6)
 Activating/deactivating
the "Dual time" function
(section 20.7)
Rev. 01 / 11.2022
RPM
t/min:s
RPM
t/min:s
RPM
t/min:s
PROG
RPM
Query:
 Hours of operation
(section 20.8)
 Number of carried-out
centrifugation runs
(section 20.8)
Setting:
 Cycle counter
(section 20.9)