Hettich EBA 280 Operating Manual page 39

Hide thumbs Also See for EBA 280:
Table of Contents

Advertisement

Message / fault
CTRL ERROR
CRC ERROR
COM ERROR
FC ERROR
61.1 – 61.21,
61.64 – 61.142
TACHO ERR
FC ERROR
FC ERROR
VERS. ERR
AB1101EN_SA
Cause
22 – 25.4
Error / defect electronics
27, 27.1
31 – 36
60,
61.22
Speed measurement error
61.23
61.153
Error / defect electronics
61.154
Invalid machine version
Rev. 01 / 11.2022
Remedy
 Perform a MAINS RESET when the
rotor has been stationary.
 The device must not be switched off
as long as the "wait" message is
being displayed. After the message
"wait ...!" is no longer displayed,
carry out a MAINS RESET.
 Perform a MAINS RESET.
 Check the loading of the rotor, see
chapter "Loading the rotor".
 Repeat the centrifugation run.
 Perform a MAINS RESET.