Activating/deactivating the "Dual time" function

In order to be able to set the runtime counter (parameter Begins at START and parameter Begins at SPEED), the
"Dual time" function must be activated. This function is activated ex works.
With the rotor at a standstill, this can be set as follows:
The procedure can be aborted at any time by pressing the
 Press and hold the button for eight
seconds.
 Press the key until the following is
displayed:
 Press the key.
START
PULSE
 Press the key until one of the following is
displayed:
Dual time : Counting the runtime.
on : Function activated.
off : Function deactivated.
 Set with the off or on keys.
 Press the key to save the setting.
START
PULSE
 Press the key once to exit the "
STOP
OPEN
Settings"
menu or press twice to exit the "
MACHINE MENU ".
AB1101EN_SA
STOP
OPEN
key. In this case, the settings are not stored.