24.3

Autoclaving

The following accessory can be autoclaved at 121°C / 250°F (20 min):

Swing-out rotors without hangers

Reductions
No statement can be made about the degree of sterility.
Autoclaving accelerates the ageing process of plastics. In addition, autoclaving may discolour plastics.
24.4

Centrifuge containers


With leakiness or after the breakage of centrifuging containers broken container parts, glass splinters and leaked
centrifugation material are to be completely removed.

The rubber inserts as well as the plastic sleeves of the rotors are to be replaced after a glass breakage.
Remaining glass splinters cause further glass breakage!

If this concerns infectious material, a disinfection process is to be executed immediately.
AB1101EN_SA
Rev. 01 / 11.2022
EN
37/45