Operating the System

X-ray Table Examination Procedures

4.8 X-ray Table Examination Procedures
4.8.1 Table Flexible Mode
4.8.1.1 General
The Table Flexible mode is designed for examinations with the detector placed on the table
top.
The overhead tube crane can be moved freely in all directions. When the height of the table
top is adjusted, the overhead tube crane is track the height of the table top in order to keep
the SID constant.
4.8.1.2 Flow of Operation
Select a Table Flexible mode examination.
The System display is show the following.
• The SID is displayed.
Activate the mode by pressing the servo button.
• The overhead tube crane moves automatically to its programmed position.
• The overhead tube crane moves to the programmed SID.
• Servo mode indication light fixed.
Adjust the table height.
• The overhead tube crane follows the Table to maintain the programmed SID.
Adjust the tube position.
• The mode is stay activated (detector holder is not follow).
• Move the imaging unit manually, if needed, to assure that the x-ray field is inside the
boundaries of the detector.
4.8.1.3 Exposure
Exposure is possible when the overhead tube crane is not moving, and the servo mode
indication light is fixed.
Note!
In "Table Flexible mode" examinations exposure is possible outside the imaging unit.
The imaging unit does not follow the overhead tube crane.