Multi Box; X-Ray Interface Unit - Canon AX-C4343W Technical Document

Digital x-ray imager 2.4 ghz/5 ghz
Table of Contents

Advertisement

3.2

Multi Box

This box connects and relays control signals between the X-ray generator, the image-
capture computer, the Docking Station and other equipment. The battery pack attached
to the detector is also charged while the wiring cable (sold separately, optional) is
connected to the detector (charging time: approx. 120 minutes).
3.3

X-ray interface unit

This unit synchronizes the timing of X-ray irradiation with the detector's capture. No
operation is required for this unit while using the system.
A
Ground terminal
B
X-ray interface cable
C
LAN cable connector
D
DC IN jack
E
POWER lamp
* Required specifications for a cable connected at installation by a service engineer are as follows:
• The X-ray interface cable must be connected to the secondary circuit of an X-ray generator that is
compliant with IEC 60601-1, and that circuit must be isolated from commercial power.
The input/output values for the X-ray generator's communication circuit must not fall outside the
specification range as indicated below:
Signal input part (SIP):
Signal output part (SOP):
• The X-ray interface box and X-ray generator must be grounded to a common protective earth.
A
B
C
Max. voltage:
Max. voltage:
Current range:
D
E
Accepts the ground.
Connects* to the X-ray generator.
Accepts the LAN cable.
Accepts the power cord of the dedicated AC
adapter.
Indicates the power status of the X-ray interface box.
5 V DC
250 V AC/30 V DC
10 mA to 2 A
3
Parts names and functions
29