Operating the System
Manual Mode
4.6.1.1 Deactivation of Manual Mode
The Manual mode is deactivated by selecting a new APR with a different auto-position.
Deactivation is also done by changing to Automatic mode .
If the same patient is examined and an APR with the same auto-position is selected, the
Manual mode is kept.
4.6.1.2 Restrictions in Manual Mode
In Manual mode the shown SID value is based on the assumption that the detector is
positioned in the same way as it should have been in an active servo mode. Note that the
SID value is not able to be calculated is some positions. No SID value is shown if the System
is not in an active servo mode when Manual mode is activated.
Tracking or other mode movements are not possible when Manual mode is active.