User Interface
Display User Interface
The mode is intended as a flexible, easy to use mode. The mode can also be used as a
parking mode.
The distance (H) shown in display for Auto position mode is the distance to the floor.
Exposure Validation
Exposure is allowed if the stand is not moving and operating properly (not in an error state).
The chosen position must have been reached successfully.
3.3.5.3 Wall Flexible Mode
General Description
The Wall Flexible mode is intended for examinations toward a wallstand.
The wallstand detector angle is controlled from the APR.
The overhead tube crane moves upon activation of the servo button to the programmed
position associated with Wall mode. The stand stops at the transport height and waits for a
change in position of the wallstand (detector height). When a change in position is detected
(the user moves/runs the wallstand up or down) the overhead tube crane moves downward
and start tracking the position of the detector.
The SID value is constantly updated as soon as the wallstand/overhead tube crane is
moved. It is possible for an operator to change the position so the SID value cannot be
calculated or would be incorrect; in those situations the display does not show a SID value.
No Wait
At the installation of the System it is possible to select, that the overhead tube crane shall not
wait for the user to move the wallstand before tracking starts. The overhead tube crane then
starts the tracking as soon as it reaches its final position.
Exposure Validation
Exposure is allowed if the stand is not moving, operating properly (not in an error state) and
the servo button is activated.