13 Appendix B

13.1 Monthly Checklist

Make a copy of this paper before filling in.
If there is any discrepancy, please use the table to make a note.
Hospital:..................................................................................................................
Id no:....................................................
Sign:.....................................................

13.1.1 Overhead Tube Crane

1. Move the overhead tube crane manually to all positions
2. Check the emergency stop. By activating the emergency
3. Choose table position and make sure the tracking is
4. Check that the SID, shown on the display of both the
5. Check the hoses for damage.
6. Check all outer cabling for damage.
7. Clean all outer surfaces, except for the lubricated column
Operator´s Manual
in X, Y and Z direction and make sure it runs smoothly
and sounds OK.
stop all motorized movements are inhibited.
activated.
Measure between the X-ray tube focal spot and the
active image receptor surface of the image receptor
holder. The measured SID shall correspond with the
displayed SID.
Move the overhead tube crane in X or Y direction and
measure between the X-ray tube focal spot and the
active detector surface of the detector holder again. The
SID is allowed to differ ±1%.
Image system and the collimator, correspond with the
measured SID.
segments. See Chapter "Cleaning" at the Instruction for
use.